Tuning with Your Analyte in LC/ESI/MS Mode
Optimizing the MS Detector Tune Automatically with Your Analyte
Thermo Electron Corporation
LTQ XL Getting Started
79
Figure 34.
Divert/Inject Valve dialog box
8. Observe the Tune Plus window and the Tune dialog box. While
automatic tuning is in progress, the LTQ XL MS detector displays
various tests in the Spectrum and Graph views in the Tune Plus window
and displays various messages in the Status group box in the Tune dialog
box. Your Tune Plus window should now look similar to the one shown
in
.
You have now successfully tuned the MS detector in ESI/MS mode for the
compound reserpine (or your analyte of interest).
Note
The most important parameters that affect the signal quality
during ESI/MS operation are electrospray voltage, ion transfer
capillary temperature, heated capillary voltage, tube lens voltage,
gases, and solution flow rate. If any one of these parameters is
changed, you need to reoptimize MS detector parameters. You can
use the Semi-Automatic tune procedure to tune the MS detector on
individual parameters.