B
High Mass Range Calibration Procedure
158
LTQ XL Getting Started
Thermo Electron Corporation
a. At the bottom of the Calibrate dialog box, click
Start
. A message
box appears with the message: Please ensure that the 500 microliter
syringe is full.
b. Verify that the syringe contains at least
420
μ
L of the high mass
range calibration solution.
c. Click OK to close the message box and return to the Calibrate
dialog box. Start the calibration.
12. Observe the Tune Plus window and the Calibrate dialog box. While the
automatic calibration is in progress the Spectrum and Graph views
display a variety of test results and the Status pane in the Calibrate
dialog box displays status messages. The system will automatically use
the average mass values for the Turbo Scan calibration. Monoisotopic
masses will automatically be used for the normal, Zoom Scan and
isolation waveform calibrations.
13. Observe the PPG2700 calibrations are complete. it may be necessary to
-the two point mass calibration.
When all calibration items are completed, your LTQ XLis properly tuned
and calibrated in the high mass range. At the conclusion of the procedure,
you must clean the syringe and API source (change the tubing and syringe,
clean the API source with acetone to remove the polypropylene glycol).