B
Enter Normal Mass Range Data into Tune Plus
Thermo Electron Corporation
LTQ XL Getting Started
151
in the Expected High Mass box. Type the Observed Low Mass and the
Observed High Mass values in their respective text boxes. These latter
values are obtained as discussed in the preceeding section:
. After entering the observed mass values click the
Execute
button.
6. In the Estimate Other Calibration Modes Using Current Full Scan box,
click
Execute
.
7. Click
Save
at the bottom right of the Manual Coarse Calibration
window.
8. Click
OK
at the bottom of the Diagnostics dialog box.