Optimizing the MS Detector with Your Analyte in APCI/MS Mode
Setting Up the MS Detector for APCI/MS Operation
120
LTQ XL Getting Started
Thermo Electron Corporation
b. Verify that the settings in your dialog box are similar to those shown
in
c. Click
OK
to close the APCI Source dialog box.
4. Define the scan parameters for tuning the MS detector in the APCI/MS
mode, as follows:
a. On the Control/Scan Mode toolbar, click
Define Scan
to open the
Define Scan dialog box.
(If your dialog box appears
different from the one shown in the figure, it is probably because the
advanced settings are not displayed. You can turn on the advanced
settings as follows: In Tune Plus, choose
ScanMode
, and then click
Advanced Scan Features
to select the option.)
b. In the Scan Description group box, in the Mass Range list box,
select
Normal
to allow for a selection of mass ranges between
m/z
150 to 2000.
c. In the Scan Rate list box, select
Normal
to specify a normal scan
rate.
d. In the Scan Type list box, select
SIM
to specify a selected ion
monitoring scan.
e. In the Scan Time group box, in the Microscans spin box, type
1
to
set the total number of microscans to 1.
f. In the Max. Inject Time spin box, type
200.000
to specify a 200 ms
maximum injection time.