Thermo Electron Corporation
LTQ XL Getting Started
115
Chapter 7
Optimizing the MS Detector
with Your Analyte in
APCI/MS Mode
This chapter provides information on optimizing the tune of your MS
detector in the APCI/MS high flow mode. It is not necessary to recalibrate
the MS detector when you switch to APCI/MS operation. You can use the
calibration settings you obtained from the successful automatic calibration
procedure you performed in the ESI/MS mode.
For APCI/MS operation you simply open a default Tune Method located in
your C:\Xcalibur\methods folder, in this case APCIhighflow.LTQTune.
From this starting point, you optimize automatically the tube lens voltage
for your particular analyte. The capillary voltage and ion transfer capillary
temperature may then be optimized manually to enhance ion transmission
This chapter includes the following sections:
•
Setting Up the Inlet for Tuning Using High-Flow Infusion
•
Setting Up the MS Detector for APCI/MS Operation
•
Optimizing the Tune of the MS Detector Automatically in APCI/MS
Mode
•
Saving the APCI/MS Tune Method
•
Cleaning the MS Detector after Tuning in APCI Mode
Note
The following procedures assume that you are familiar with your
LTQ XL instrument and the Tune Plus window. For more information,
refer to the LTQ XL online Help,
LTQ XL Getting Connected,
or the
LTQ
XL Hardware Manual
.
Ensure that you have completed the procedures in the topics Tuning and
Calibrating Automatically in the ESI/MS Mode and Setting Up to
Acquire Data in the APCI/MS Mode.