3
Tuning and Calibrating Automatically in the ESI/MS Mode
Calibrating the MS Detector Automatically
62
LTQ XL Getting Started
Thermo Electron Corporation
Calibrating the MS
Detector
Automatically
Calibrate the MS detector automatically from the Tune Plus window
1. Choose
Control > Calibrate
to display the Calibrate dialog box.
2. If necessary, click the
Automatic
tab to display the Automatic
calibration page. See
Figure 28.
Calibrate dialog box, showing the Automatic calibration page
3. Start the automatic calibration procedure, as follows:
a. Click
Start
. A message box displays the following message:
Please ensure that the 500 microliter syringe is full.