4
Tuning with Your Analyte in LC/ESI/MS Mode
70
LTQ XL Getting Started
Thermo Electron Corporation
The capillary is heated to maximize the ion transmission to the MS detector.
For ESI only, you set the ion transfer capillary temperature proportional to
the flow rate of your solution. Refer to Table 1-2 for guidelines for setting
operating parameters for LC/ESI/MS. For this procedure, the ion transfer
capillary temperature is set to 350 °C, and the sheath gas is set to 30.
To tune the MS detector in the ESI/MS (high-flow) mode using your
analyte, perform the following tasks:
• Set up the MS detector for your specific analyte from the Tune Plus
window.
• Infuse your analyte into the MS detector using a syringe pump
connected to the LC with a Tee union.
• Optimize the MS detector parameters for your analyte while the
solution flows into the MS detector.
This chapter contains the following topics:
•
Setting Up to Introduce Sample by Syringe Pump into Solvent Flow
from an LC
•
Setting Up to Tune the MS Detector with Your Analyte
•
Optimizing the MS Detector Tune Automatically with Your Analyte
•
Note
1. If your experiment is performed at a flow rate below 10
μ
L/min, and
the results you want can be obtained without optimizing the MS
detector on your particular analyte, go to
Sample Data Using the Tune Plus Window”
to acquire sample data.
2. Before you optimize the tune for your analyte of interest, ensure that
the LTQ XL MS detector has been calibrated within the previous
three months. If the system needs to be calibrated, refer to the
procedures in