3
Tuning and Calibrating Automatically in the ESI/MS Mode
Setting Up the MS Detector in the Xcalibur Data System for Tuning and Calibration
Thermo Electron Corporation
LTQ XL Getting Started
51
Figure 20.
Define Scan dialog box, showing the default settings for ESI/MS operation
i.
In the Scan Ranges group box, in the Scan Ranges table, in the First
Mass text box, enter
150
to set the first mass for the scan range to
m/z
150.
j.
In the Last Mass text box, enter
2000
to set the last mass for the scan
range to
m/z
2000.
k. Ensure that the settings in your Define Scan dialog box are the same
as those shown in
.
l.
Click
OK
to apply the MS detector scan parameters and to close the
Define Scan dialog box.
6. On the Control/Scan Mode toolbar, click the
Centroid/Profile
button
to toggle the data type to profile. (The picture on the button should be
the same as that shown here.)
7. Click the
Positive/Negative
button to toggle the ion polarity mode to
positive. (The picture on the button should be the same as that shown
here).
The MS detector is now properly set up in the Xcalibur data system for
tuning and calibration in the ESI/MS mode.