Acquiring ESI Sample Data Using the Tune Plus Window
Setting Up to Acquire MS/MS Data in the Full Scan Type
88
LTQ XL Getting Started
Thermo Electron Corporation
10. When you have clearly identified a product ion mass-to-charge ratio for
reserpine (or your analyte of interest), click the
Tune
button to display
the Tune dialog box.
11. In the Tune dialog box, click the
Collision Energy
tab to display the
page. See
.
Figure 39.
Define Scan dialog box, showing typical settings for acquiring MS/MS data in the Full scan type on reserpine
12. Click the
Product Ion Mass
option button to make active the spin box.
Type
397.2
to specify the product ion at
m/z
397.2 for reserpine. The
LTQ XL MS detector can optimize collision energy automatically by
using this product ion of reserpine.