6
Setting Up the Ion Source for Acquiring Data in APCI/MS/MS Mode
Installing the APCI Probe
112
LTQ XL Getting Started
Thermo Electron Corporation
Figure 55.
APCI probe
7. Lock the probe in place by rotating the probe locking lever towards the
front of the housing; closing the probe locking lever towards the rear of
the ion source housing might make it difficult to unlock.
8. Unplug the vaporizer heater cable from the ESI interlock plug on the
ion source housing.
9. Connect the vaporizer heater cable to the vaporizer heater cable socket
on the APCI probe.
10. Connect the sheath gas fitting (blue) to the sheath gas inlet (S) on the
probe. (See Figure 55)
11. Connect the auxiliary gas fitting (green) to the auxiliary gas inlet (A) on
the probe. (Figure 55)
12. Connect the sample transfer line to the APCI probe inlet.
The APCI probe is now properly installed in the Ion Max ion source
housing.
Vaporizer Heater
Cable Socket
Guide Pin
Sheath Gas
Inlet (S)
Auxilary
Gas Inlet (A)
Sample Inlet
Fitting