Tuning with Your Analyte in LC/ESI/MS Mode
Saving the ESI/MS Tune Method
Thermo Electron Corporation
LTQ XL Getting Started
81
Saving the ESI/MS
Tune Method
Save your ESI/MS Tune Method (for a high-flow experiment using your
analyte) when automatic tuning is complete
1. Choose
File > Save As
to display the Save As dialog box. See
.
Figure 36.
Save As dialog box, showing files in the folder C:\Xcalibur\methods
2. Select the C:\Xcalibur\methods folder.
3. Click the
File Name
text box, and enter
reserpine
(or the name of your
analyte of interest).
4. Click
Save
to save the Tune Method, and return to the Tune Plus
window. Note that the Tune Method is named reserpine.LTQTune.
Note
Save the Tune Method while the MS detector is On, if any of the
ion source parameters are different from those with which you started.