2
Setting Up the Ion Source for Tuning and Calibrating the MS Detector
Placing the LC/MS System in Standby
28
LTQ XL Getting Started
Thermo Electron Corporation
Placing the LC/MS
System in Standby
The LC/MS system needs to be placed in Standby condition before you can
remove the ion source.
Place the LC/MS system in Standby
1. If necessary, stop the flow of solvent to the API source as follows:
a. If the Xcalibur data system is not already open, choose
Start >
Programs > Xcalibur > Xcalibur
from the Windows
®
taskbar to
open the Xcalibur window.
b. In the Xcalibur Home Page window – Roadmap view, choose
GoTo > Instrument Setup
to open the Instrument Setup window.
c. Click the
Surveyor® MS Pump
button on the view bar in the
Instrument Setup window to display the Surveyor MS Pump view.
d. Choose
Surveyor MS Pump > Direct Control
to open the Surveyor
MS Pump Direct Control dialog box.
e. In the Direct Control dialog box, click the
Pump Off
button to stop
the MS pump.
2. If Tune Plus is not already open, choose
Start > Programs > Xcalibur >
LTQ XLTune
from the taskbar to open Tune Plus.
You can determine the state of the MS detector by observing the state of
the On/Standby button on the Control / Scan Mode toolbar. (The three
different states of the On/Standby button are shown at the left.)
3. If the MS detector is On, click the
On/Standby
button to place the MS
detector in the Standby mode. When the MS detector is in Standby, the
LTQ XL MS detector turns off the ion source sheath gas, auxiliary gas,
and high voltage.
The LC/MS system is now in Standby and it is safe to remove the ion
source.
If the ESI probe is already installed in the Ion Max™ ion source housing,
leave the LC/MS system in Standby and go to
Calibrating Automatically in the ESI/MS Mode”
If the ESI probe is not already installed in the Ion Max ion source housing,
go to the next section,
.
On
Off
Standby