Setting Up the Ion Source for Tuning and Calibrating the MS Detector
Removing the APCI Probe
Thermo Electron Corporation
LTQ XL Getting Started
29
Removing the APCI
Probe
This topic describes how to remove the APCI probe from the Ion Max ion
source housing.
Remove the APCI probe
1. Unplug the vaporizer heater cable from the vaporizer heater cable socket
on the APCI probe. See Figure 5.
2. Disconnect the sample transfer line from the APCI probe. (See
3. Remove the auxiliary gas line (green-colored fitting) from the APCI
4. Remove the sheath gas line (blue-colored fitting) from the APCI probe.
Note
The following procedures assume that you are familiar with your
instrument and software. If you need additional guidance, refer to LTQ
XL online Help,
LTQ XL Getting Connected
,
Ion Max API Source
Hardware Manual
, or the
LTQ XL Hardware Manual
.
CAUTION
AVOID BURNS.
At operating temperatures, the APCI
vaporizer can severely burn you! The APCI vaporizer typically operates
between 400 and 600
°
C.
Always allow the heated vaporizer to cool to
room temperature (for approximately 20 min) before you touch or
remove this component.