Setting Up the Ion Source for Tuning and Calibrating the MS Detector
Removing the APCI Probe
30
LTQ XL Getting Started
Thermo Electron Corporation
Figure 5.
Ion Max ion source housing with APCI probe installed
5. Remove the APCI probe as follows:
a. Connect the vaporizer heater cable to the ESI interlock socket on
the ion source housing. See
b. Release the probe locking lever to loosen the probe collar. You might
need to unscrew the lever a few turns to permit probe movement.
c. Carefully pull the probe straight back in the port in the housing
until it meets with the slot in the API interlock block. The guide pin
on the probe manifold will prevent you from rotating the probe
until the pin is aligned with the slot in the API interlock block.
Once the probe is all the way back and aligned with the slot, turn
the probe 45 degrees counter-clockwise to free the probe from the
alignment notch.
d. Pull the probe straight out to remove it from the ion source housing.
e. Store the APCI probe in its original shipping container.
Vaporizer
Heater
Cable
Auxiliary
Gas
Fitting
Sheath
Gas
Fitting
Corona
Needle
High
Voltage
Receptacle
8 kV
Cable
Sample
Transfer
Line
CAUTION
AVOID BURNS.
At operating temperatures, the APCI
vaporizer can severely burn you! The APCI vaporizer typically operates
between 400 and 600
°
C.
Always allow the heated vaporizer to cool to
room temperature (for approximately 20 min) before you touch or
remove this component.