3
Tuning and Calibrating Automatically in the ESI/MS Mode
Testing the Operation of the MS Detector in the ESI/MS Mode
52
LTQ XL Getting Started
Thermo Electron Corporation
Testing the Operation
of the MS Detector in
the ESI/MS Mode
You are now ready to test whether your MS detector is operating properly.
To test for proper operation, you infuse the calibration solution into the ESI
source, and then you monitor the real-time display of the mass spectrum of
calibration solution. You want to ensure that a stable spray of solution enters
the MS detector.
Test the operation of the MS detector in the ESI/MS mode
1. Click the
Syringe Pump
button to display the Syringe Pump dialog
box. See
.
Figure 21.
Syringe Pump dialog box
2. Turn on the syringe pump and set an infusion flow rate of 5
μ
L/min, as
follows:
a. In the Flow Control group box, click the
On
option button to make
active the Flow Rate spin box.
b. Type
5
in the Flow Rate spin box to specify a rate of 5
μ
L/min.
c. If you are using a standard Unimetrics (or Hamilton) syringe, set up
the syringe parameters as follows:
Note
This procedure assumes that you are using the 500-
μ
L
Unimetrics syringe that is provided with your LTQ XL system. If
you are using another type of syringe, select the option button
corresponding to your syringe.