7
Maintenance and Service
Page 198
ISQ EC and ISQ EM Mass Spectrometer Operating Manual
16)
Press on the electrical pin through the hole where the first jumper
wire was just unplugged to remove the skimmer cone. All that holds
the skimmer cone in place are the Q00 assembly and its electrical
connection, so if you unplugged the wire in the previous step, it
should just take a little bit of force to remove it. See Figure 119.
Figure 119: Removing the Skimmer Cone
17)
Remove the three screws holding the tube lens in place. To remove
the tube lens, press on the electrical pin through the second hole
where the jumper was unplugged using a tool like an Allen wrench.
See Figure 120.
Figure 120: Removing Tube Lens Screws
18)
Now you should have removed all of the optics for cleaning. See
Figure 121: Optics parts.
Содержание ISQ EC
Страница 1: ...ISQ EC and ISQ EM Mass Spectrometer Operating Manual 1R120591 0002 Revision E February 2019 ...
Страница 4: ...Page 4 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 66: ...5 Installation Page 66 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 135: ...6 Operation ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 135 ...
Страница 163: ...6 Operation ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 163 Figure 90 Venting the mass spectrometer ...
Страница 166: ...6 Operation Page 166 ISQ EC and ISQ EM Mass Spectrometer Operating Manual MS LC IC Start port Integrion RLY port ...
Страница 204: ...7 Maintenance and Service Page 204 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 228: ...12 Index Page 228 ISQ EC and ISQ EM Mass Spectrometer Operating Manual 12 Index ...
Страница 229: ...12 Index ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 229 ...
Страница 232: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...