6
Operation
ISQ EC and ISQ EM Mass Spectrometer Operating Manual
Page 157
6.7
Shutting Down the MS Instrument
If the instrument will not be operated for some time, follow the instructions
on enabling standby for detector in this section.
6.7.1
Short-Term Standby
If the instrument is unused for a few days, allow it to remain connected to
the system, and for the system to remain powered on.
To interrupt operation of the instrument for a short period (short-term
shutdown), for example, overnight, observe these guidelines for the
detector:
Keep the instrument connected to the system flow path.
Turn off the pump flow.
If you operate the instrument from a chromatography data system,
keep the detector connected to the data system.
When resuming operation, let the flow equilibrate and verify that the
operating parameters for the other system modules are set as required
before proceeding.
To set Smart Standby parameters, go to the Chromatography Studio and
select Smart Standby. Set the standby parameters for the MS using the
check boxes. See
Содержание ISQ EC
Страница 1: ...ISQ EC and ISQ EM Mass Spectrometer Operating Manual 1R120591 0002 Revision E February 2019 ...
Страница 4: ...Page 4 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 66: ...5 Installation Page 66 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 135: ...6 Operation ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 135 ...
Страница 163: ...6 Operation ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 163 Figure 90 Venting the mass spectrometer ...
Страница 166: ...6 Operation Page 166 ISQ EC and ISQ EM Mass Spectrometer Operating Manual MS LC IC Start port Integrion RLY port ...
Страница 204: ...7 Maintenance and Service Page 204 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 228: ...12 Index Page 228 ISQ EC and ISQ EM Mass Spectrometer Operating Manual 12 Index ...
Страница 229: ...12 Index ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 229 ...
Страница 232: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...