7.3
General Rules for Maintenance and Service
For successful maintenance and service procedures, follow these rules and
recommendations:
Before starting maintenance or service procedures, shut down the
instrument when instructed to do so.
Use only the replacement parts specifically authorized and qualified for
the detector by Thermo Fisher Scientific. For ordering information, see
Section 10.3
Follow all instructions systematically and use the tools recommended
for the procedure.
Before opening the flow path to replace capillaries in the system, turn
off the pump flow and wait until the system pressure is down to zero.
Dirty components can contaminate the system. Contamination leads to
poor performance of the modules and entire system or can even cause
damage to the modules and system. Therefore:
Always wear appropriate gloves.
Place the components only on a clean, lint-free surface.
Keep your tools clean.
Use only lint-free cloth for cleaning.
When disconnecting capillaries, protect open flow connections with
plugs.
Содержание ISQ EC
Страница 1: ...ISQ EC and ISQ EM Mass Spectrometer Operating Manual 1R120591 0002 Revision E February 2019 ...
Страница 4: ...Page 4 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 66: ...5 Installation Page 66 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 135: ...6 Operation ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 135 ...
Страница 163: ...6 Operation ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 163 Figure 90 Venting the mass spectrometer ...
Страница 166: ...6 Operation Page 166 ISQ EC and ISQ EM Mass Spectrometer Operating Manual MS LC IC Start port Integrion RLY port ...
Страница 204: ...7 Maintenance and Service Page 204 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 228: ...12 Index Page 228 ISQ EC and ISQ EM Mass Spectrometer Operating Manual 12 Index ...
Страница 229: ...12 Index ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 229 ...
Страница 232: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...