6
Operation
Page 148
ISQ EC and ISQ EM Mass Spectrometer Operating Manual
Table 11: APCI source parameters: Advanced view
6.6.4
Method Editor—Scan Settings,
The basic mode scan table is designed to be minimalistic for easy method
creation. To enter a full scan or selected ion monitoring (SIM) scan, simply
enter the desired mass of interest and the preferred polarity. The scan name
and Source CID values are optional (defaults to 0.0 V). Scan and dwell times
are automatically calculated based upon the inputted acquisition rate. Enter
the minimum peak width value and scans per peak desired, and the
sampling rate will be calculated in Hz. If the minimum peak width value is
unknown, adjust the values until the desired rate is selected. The sampling
rate is applied to each scan line item entered in the scan table. See Figure
82.
Parameter
Description
Vaporizer temperature
Heat applied the auxiliary gas for desolvation. 0 °C – 550 °C
Ion transfer tube
temperature
Heat applied to the transfer tube at the source to aid in desolvation
and transmission. 150 °C – 400 °C
Sheath gas pressure
Primary gas flow for liquid nebulization. 0 psig – 80 psig
Aux gas pressure
Auxiliary heated gas that aids in desolvation. 0 psig – 15 psig
(Recommended range: 0 – 10 psig)
Sweep gas pressure
Counter-current gas flow at the sweep cone to improve robustness.
0 psig – 2 psig
Source current positive
ions
APCI corona discharge needle current for positive ionization. 0 µA –
50 µA (Recommended range: 0 – 8 µA)
Source current negative
ions
APCI corona discharge needle current for negative ionization. 0 µA –
50 µA (Recommended range: 0 – 8 µA)
Содержание ISQ EC
Страница 1: ...ISQ EC and ISQ EM Mass Spectrometer Operating Manual 1R120591 0002 Revision E February 2019 ...
Страница 4: ...Page 4 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 66: ...5 Installation Page 66 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 135: ...6 Operation ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 135 ...
Страница 163: ...6 Operation ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 163 Figure 90 Venting the mass spectrometer ...
Страница 166: ...6 Operation Page 166 ISQ EC and ISQ EM Mass Spectrometer Operating Manual MS LC IC Start port Integrion RLY port ...
Страница 204: ...7 Maintenance and Service Page 204 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 228: ...12 Index Page 228 ISQ EC and ISQ EM Mass Spectrometer Operating Manual 12 Index ...
Страница 229: ...12 Index ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 229 ...
Страница 232: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...