ISQ EC and ISQ EM Mass Spectrometer Operating Manual
TIP—Once the Play button has been pressed, the instrument settings can be
changed in real-time.
Parameter
Description
Play button
Initiates real time scanning
Stop button
Ends real-time scanning
Real time data
display
Shows the real-time data acquisition once the Play button has been
selected. The X (m/z mass range) and Y (relative signal intensity)
are displayed.
Status window
Displays the current source type and instrument status
Scan type
User selectable for full scan and SIM modes of operation
Spectrum Type Select the spectrum type: profile or centroid. . For SIM methods,
select profile mode only.
Polarity
Select the probe polarity
Use calibrant
A check in this box will actuate the internal valve and initiate an
internal calibrant infusion into the MS instrument
Use tune value
Use to enter a custom value for the tube lens
Ion transfer
tube temp
Select the heat applied to transfer tube after the exit cone
Vaporizer temp Select the heat applied to the ESI probe out of the manifold for
aiding in desolvation
Source voltage
Select the voltage applied to the ESI capillary for ionization
Source CID
voltage
Voltage applied within the source to induce collisional dissociation
Tube lens
voltage
Select the voltage applied to the tube lens
Sheath gas
pressure
Select the main nitrogen gas pressure used for desolvation /
nebulization.
Aux gas
pressure
Select the gas flow around the probe outer manifold
Sweep gas
pressure
Select the counter-flow gas pressure coming out of sweep cone
Load calibration
masses
Button that automatically provides pre-selected full-scan masses
for internal calibrant.
Mass Table
Enter selected mass and/or mass range of interest. Real-time data
display will auto update with selection
Table 7
: Real time scanning list of features and descriptions
Entering analyte masses of interest into the Mass Table will automatically
update the real time data display for the given mass or mass range of
interest.
Содержание ISQ EC
Страница 1: ...ISQ EC and ISQ EM Mass Spectrometer Operating Manual 1R120591 0002 Revision E February 2019 ...
Страница 4: ...Page 4 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 66: ...5 Installation Page 66 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 135: ...6 Operation ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 135 ...
Страница 163: ...6 Operation ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 163 Figure 90 Venting the mass spectrometer ...
Страница 166: ...6 Operation Page 166 ISQ EC and ISQ EM Mass Spectrometer Operating Manual MS LC IC Start port Integrion RLY port ...
Страница 204: ...7 Maintenance and Service Page 204 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 228: ...12 Index Page 228 ISQ EC and ISQ EM Mass Spectrometer Operating Manual 12 Index ...
Страница 229: ...12 Index ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 229 ...
Страница 232: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...