Maintenance and Service
ISQ EC and ISQ EM Mass Spectrometer Operating Manual
Page 191
Figure 107: Removing source housing screws
5)
Use a 2.5 mm hex key to gently screw in the three set screws as
shown in Figure 108: Screwing in the set screws. The set screws will
push the cone assembly out of the instrument
Figure 108: Screwing in the set screws
6)
The cone should move out enough to break the internal seal. Once
that occurs, remove the cone and front optics by hand. See Figure
109: View of front optics with seal broken