Figure 81: MS instrument method editor – Source settings, advanced view
Additional features of the advanced source settings are shown in Table 10
below.
Table 10: HESI source parameters: Advanced view
Parameter
Description
Vaporizer temperature
Heat applied the auxiliary gas for desolvation. 0 °C – 550 °C
Ion transfer tube
temperature
Heat applied to the transfer tube at the source to aid in desolvation
and transmission. 150 °C – 400 °C
Sheath gas pressure
Primary gas flow for liquid nebulization. 0 psig – 80 psig
Aux gas pressure
Auxiliary heated gas that aids in desolvation. 0 psig – 15 psig
Sweep gas pressure
Counter-current gas flow at the sweep cone to improve robustness.
0 psig – 2 psig
Source voltage positive
ions
HESI capillary voltage for positive ionization. 0 V – 8000 V
Source voltage negative
ions
HESI capillary voltage for negative ionization -8000V – 0 V
Содержание ISQ EC
Страница 1: ...ISQ EC and ISQ EM Mass Spectrometer Operating Manual 1R120591 0002 Revision E February 2019 ...
Страница 4: ...Page 4 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 66: ...5 Installation Page 66 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 135: ...6 Operation ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 135 ...
Страница 163: ...6 Operation ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 163 Figure 90 Venting the mass spectrometer ...
Страница 166: ...6 Operation Page 166 ISQ EC and ISQ EM Mass Spectrometer Operating Manual MS LC IC Start port Integrion RLY port ...
Страница 204: ...7 Maintenance and Service Page 204 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 228: ...12 Index Page 228 ISQ EC and ISQ EM Mass Spectrometer Operating Manual 12 Index ...
Страница 229: ...12 Index ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 229 ...
Страница 232: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...