3
Instrument Overview
ISQ EC and ISQ EM Mass Spectrometer Operating Manual
Page 43
3.3
Software Operation
The mass spectromter is designed to be operated from a computer
configured with Chromeleon software. Chromeleon software provides
complete instrument control, data acquisition, and data management.
The data system serves as the user interface to the mass spectrometer,
autosampler, LC pump, and syringe pump. Refer to the Chromeleon Help for
more information about the ISQ EC and ISQ EM Single Quadrupole Mass
Spectrometer data processing and instrument control. Each sequence of
loading a mass analyzer with ions followed by mass analysis of the ions is
called a scan. The mass spectrometer utilizes both a full scan type and/or
selected ion monitoring (SIM) scan type to create, fragment, and detect
ions. The ability to vary not only the ionization and ion polarity modes, but
the scan mode and scan type, provides greater flexibility in the
instrumentation for solving complex analytical problems.
The method editor in Chromeleon allows the user to set up scans in a
variety of ways, combining both full scans and SIM scans. The user can also
call for scans of either or both polarities. Simplified ion source controls and
a simplified method editor make setting up an instrument method easier
than ever, while the ability to switch to manual controls means that users
who wish to control each device independently do not lose that freedom.
For more information on the method editor, please see the help file in the
Chromeleon data system. Users can set up full scans to get a picture of the
entire mass spectrum as well as set up SIM scans to monitor for the
presence of specific ions. The ability to apply source collision induced
dissociation (sCID) gives the user the capability to reduce the presence of
adducts, or to create confirmation ions via fragmentation in the ion optics of
the system.
Содержание ISQ EC
Страница 1: ...ISQ EC and ISQ EM Mass Spectrometer Operating Manual 1R120591 0002 Revision E February 2019 ...
Страница 4: ...Page 4 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 66: ...5 Installation Page 66 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 135: ...6 Operation ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 135 ...
Страница 163: ...6 Operation ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 163 Figure 90 Venting the mass spectrometer ...
Страница 166: ...6 Operation Page 166 ISQ EC and ISQ EM Mass Spectrometer Operating Manual MS LC IC Start port Integrion RLY port ...
Страница 204: ...7 Maintenance and Service Page 204 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 228: ...12 Index Page 228 ISQ EC and ISQ EM Mass Spectrometer Operating Manual 12 Index ...
Страница 229: ...12 Index ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 229 ...
Страница 232: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...