ISQ EC and ISQ EM Mass Spectrometer Operating Manual
Table 15: Scan mode parameters: Scans Pane
Table 16: Scan mode parameters: Groups Pane
Table 17: Scan mode parameters: Fraction Collection Channels Pane
6.6.7
Method Development
Often, a method needs to be optimized for enhanced sensitivity or
improved qualitative detection. Whether starting from a fresh method or
transferring a protocol from a different system, it is important to take a
diligent approach to obtaining the proper source and scan parameters.
Some things to consider: mass-to-charge ratio, polarity, adduct formation,
mobile phase composition, mobile phase flow rate, etc. It is always
important to verify the m/z being detected versus assuming an [M+H]
+
or
[M-H]
-
ion as there are many possible adduct formations. Polarity is another
important consideration as some molecules favor protonation or de-
Source CID voltage
Accelerates ions into the background gas to aid in desolvation. Can
increase sensitivity or reduce noise for some analytes. Range: 0-100
V, default: 0 V (no source CID)
Tube lens voltage
Voltage applied to tube lens after the ion transfer tube. 0 V – 300 V.
Higher values can aid in desolvation. Lower values can reduce
fragmentation. Can increase sensitivity or reduce noise for some
analytes.
Note: Use 0 to 300 V for positive ions. Use 0 to -300 V for negative
ions. Use higher values for higher masses. If Last Tune is input, the
tube lens will be run with the values stored in the tune file. Last
Tune is the standard way to run the tube lens.
Parameter
Description
Time (min)
Time that the MS begins to acquire this particular group of scans
Total scan time (sec)
Sum of all the scans in each segment
Chrom. filter peak width
(sec)
Sets the peak width to match the width of the HPLC or IC peak
Detector gain
Sets the detector gain to the one in the previous tune file or may be
set manually
Parameter
Description
Name
The name of the channel for fraction collection
Scan Filter
Based on the scan information entered into the Scans pane
Target mass (min)
The minimum mass scanned within the scans range
Target mass (max)
The maximum mass scanned within the scans range
Содержание ISQ EC
Страница 1: ...ISQ EC and ISQ EM Mass Spectrometer Operating Manual 1R120591 0002 Revision E February 2019 ...
Страница 4: ...Page 4 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 66: ...5 Installation Page 66 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 135: ...6 Operation ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 135 ...
Страница 163: ...6 Operation ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 163 Figure 90 Venting the mass spectrometer ...
Страница 166: ...6 Operation Page 166 ISQ EC and ISQ EM Mass Spectrometer Operating Manual MS LC IC Start port Integrion RLY port ...
Страница 204: ...7 Maintenance and Service Page 204 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 228: ...12 Index Page 228 ISQ EC and ISQ EM Mass Spectrometer Operating Manual 12 Index ...
Страница 229: ...12 Index ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 229 ...
Страница 232: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...