8
Troubleshooting
Page 208
ISQ EC and ISQ EM Mass Spectrometer Operating Manual
8.2
Troubleshooting Problems with Autotune
Use Table 20 below to diagnose and solve problems with Autotune.
Symptom
Cause
Autotune fails at beginning
Chromatograph not set to 50 µL/min. If there is no liquid flow from the
chromatograph to push the calibrant into the ion source, the autotune will fail.
Autotune runs for a while
but then fails
Chromatograph not set to 50 µL/min. If the calibrant is pushed to the ion source
more quickly, the loop will run out before the instrument refills it. If it is pushed too
slowly, the electrospray may not be stable.
Autotune runs for a while
but then fails
If the system is failing autotune with a message that the spray stability check has
failed, check the liquid connections for leaks.
Autotune runs for a while
but then fails
An incorrectly adjusted electrospray needle. Adjust electrospray needle.
Autotune runs for a while
but then fails
The flow rate out of the liquid pump is unstable. AXP pumps in particular require
high back pressures to provide a stable flow. When using an AXP pump,
backpressure coils should be installed between the pump and the instrument to
allow the pump to run at a stable flow rate.
Reference solution not
flowing down the drain
during fill time
Out of calibrant. Replace the calibrant bottle.
Reference solution not
flowing down the drain
during fill time
Green PEEK tube in the bottle may be above the liquid level. Adjust the position of
the tubing so it is near the bottom of the bottle. It should not touch the bottom or
it may pick up particulates and clog the tubing.
Reference solution not
flowing down the drain
during fill time
No nitrogen to the instrument
Reference solution not
flowing down the drain or
flowing too slowly during fill
time
The clear plastic pressurization line, green PEEK, or calibrant loop may be kinked.
Alternatively, the ferrules on the green PEEK tubing may be overtightened.
Содержание ISQ EC
Страница 1: ...ISQ EC and ISQ EM Mass Spectrometer Operating Manual 1R120591 0002 Revision E February 2019 ...
Страница 4: ...Page 4 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 66: ...5 Installation Page 66 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 135: ...6 Operation ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 135 ...
Страница 163: ...6 Operation ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 163 Figure 90 Venting the mass spectrometer ...
Страница 166: ...6 Operation Page 166 ISQ EC and ISQ EM Mass Spectrometer Operating Manual MS LC IC Start port Integrion RLY port ...
Страница 204: ...7 Maintenance and Service Page 204 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 228: ...12 Index Page 228 ISQ EC and ISQ EM Mass Spectrometer Operating Manual 12 Index ...
Страница 229: ...12 Index ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 229 ...
Страница 232: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...