8
Troubleshooting
ISQ EC and ISQ EM Mass Spectrometer Operating Manual
Page 209
Symptom
Cause
Reference solution not
flowing down the drain or
flowing too slowly during fill
time
The small orifice on the top of the calibrant bottle may be clogged.
Reference solution does not
stop flowing after fill time
The bottle is not depressurizing. The frit on the valve of the calibrant delivery
system may be clogged. This valve is located on the gas control module under the
main power supply inside of the instrument.
Reference solution does not
stop flowing after fill time
The waste line may be inserted too far down the drain tube. This will cause the
liquid to be siphoned from the bottle.
Table 20
: Autotune problems
TIP If you are unable to resolve a problem following the instructions given
here or if you experience problems that are not covered in this section,
contact Thermo Fisher Scientific Technical Support for assistance.
Содержание ISQ EC
Страница 1: ...ISQ EC and ISQ EM Mass Spectrometer Operating Manual 1R120591 0002 Revision E February 2019 ...
Страница 4: ...Page 4 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 66: ...5 Installation Page 66 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 135: ...6 Operation ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 135 ...
Страница 163: ...6 Operation ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 163 Figure 90 Venting the mass spectrometer ...
Страница 166: ...6 Operation Page 166 ISQ EC and ISQ EM Mass Spectrometer Operating Manual MS LC IC Start port Integrion RLY port ...
Страница 204: ...7 Maintenance and Service Page 204 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 228: ...12 Index Page 228 ISQ EC and ISQ EM Mass Spectrometer Operating Manual 12 Index ...
Страница 229: ...12 Index ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 229 ...
Страница 232: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...