7.4.6
Replacing the APCI Needle
Follow the steps below to replace the APCI needle.
1)
Follow the instructions for removing the APCI probe in Section 7.4.5.
2)
Ensure the APCI needle is down so that it can be accessed. See Figure
105.
Figure 105: Locating the APCI Needle
3)
With gloved hands, pull the APCI needle out of the source housing and
install a new needle. See Figure 106.