Figure 96: Aligning the pin on the spray probe with the source housing
17)
Screw the probe into the source housing and reconnect the liquid
line.
TIP There can be some variation between electrospray needles and spray
probes. A ¼ turn in or out from the pictured penetration could be correct,
depending on the tolerances of the probe itself. If the beam is unstable, try
both directions.
7.4.5
Replacing the APCI Probe Capillaries for the ISQ EM
Follow the instructions in this section to replace the APCI probe capillaries.
CAUTION—Eye Damage
Please be sure to wear protective eyewear during the probe
replacement procedure due to exposure to elevated levels of nitrogen
gas originating from the source housing.
Содержание ISQ EC
Страница 1: ...ISQ EC and ISQ EM Mass Spectrometer Operating Manual 1R120591 0002 Revision E February 2019 ...
Страница 4: ...Page 4 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 66: ...5 Installation Page 66 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 135: ...6 Operation ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 135 ...
Страница 163: ...6 Operation ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 163 Figure 90 Venting the mass spectrometer ...
Страница 166: ...6 Operation Page 166 ISQ EC and ISQ EM Mass Spectrometer Operating Manual MS LC IC Start port Integrion RLY port ...
Страница 204: ...7 Maintenance and Service Page 204 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 228: ...12 Index Page 228 ISQ EC and ISQ EM Mass Spectrometer Operating Manual 12 Index ...
Страница 229: ...12 Index ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 229 ...
Страница 232: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...