Maintenance and Service
ISQ EC and ISQ EM Mass Spectrometer Operating Manual
Page 199
Figure 121: Optics parts
19)
To clean the optics:
a.
Using water and Liquinox, scrub the optics using a soft toothbrush.
b.
Sonicate the optics in water and Liquinox for 5 to 10 minutes.
c.
Rinse the optics parts using 18 MΩ or Optima LCMS grade water.
d.
Sonicate the optics parts in 18 MΩ or Optima LCMS grade water for
5 to 10 minutes.
e.
Rinse the optics parts using Optima LCMS grade methanol.
f.
Sonicate the optics parts in Optima LCMS grade methanol.
g.
Blow the optics parts dry with nitrogen gas.
h.
Reverse the steps above to reassemble the optics and reinstall the
assembly into the MS.
Содержание ISQ EC
Страница 1: ...ISQ EC and ISQ EM Mass Spectrometer Operating Manual 1R120591 0002 Revision E February 2019 ...
Страница 4: ...Page 4 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 66: ...5 Installation Page 66 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 135: ...6 Operation ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 135 ...
Страница 163: ...6 Operation ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 163 Figure 90 Venting the mass spectrometer ...
Страница 166: ...6 Operation Page 166 ISQ EC and ISQ EM Mass Spectrometer Operating Manual MS LC IC Start port Integrion RLY port ...
Страница 204: ...7 Maintenance and Service Page 204 ISQ EC and ISQ EM Mass Spectrometer Operating Manual ...
Страница 228: ...12 Index Page 228 ISQ EC and ISQ EM Mass Spectrometer Operating Manual 12 Index ...
Страница 229: ...12 Index ISQ EC and ISQ EM Mass Spectrometer Operating Manual Page 229 ...
Страница 232: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...