Operation
Settings
3-52
42iQHL Instruction Manual
Thermo Scientific
●
Chamber Temp Sensor Open:
Displays OK/Fail for chamber
temperature sensor open.
●
Cooler Temperature:
Displays OK/Fail for cooler temperature. User
can adjust low and high alarm limits.
●
Cooler Temp Sensor Short:
Displays OK/Fail for cooler temperature
sensor short.
●
Cooler Temp Sensor Open:
Displays OK/Fail for cooler temperature
sensor open.
●
Cooler Current:
Displays the cooler current reading.
●
Cooler Current too Low:
Displays OK/Fail for cooler current too
low.
●
Cooler Current too High:
Displays OK/Fail for cooler current too
high.
●
Cooler Voltage:
Displays the current cooler voltage reading.
●
Cooler Voltage too Low:
Displays OK/Fail for cooler voltage too low.
●
Cooler Voltage too High:
Displays OK/Fail for cooler voltage too
high.
●
PMT Voltage:
Displays the current PMT voltage reading.
●
PMT Voltage too Low:
Displays OK/Fail for PMT voltage too low.
●
PMT Voltage too High:
Displays OK/Fail for PMT voltage too
high.
●
Frequency:
Displays the current frequency reading.
●
Frequency too Low:
Displays OK/Fail for frequency too low.
●
Frequency too High:
Displays OK/Fail for frequency too high.
●
Low Gain:
Displays the current low gain reading.
●
Gain:
Displays the current gain reading.
●
Board Communication:
Displays OK/Fail for board communication
status.
●
Power Supply:
Displays OK/Fail of power supplies. Power supply
goes red if any voltages are outside their limits. No voltage rows
ever get highlighted.
●
3.3 V Diagnostic:
Displays current voltage readings. Alarm
limits are not changeable.
●
5 V Diagnostic:
Displays current voltage readings. Alarm limits
are not changeable.
Содержание 42iQHL
Страница 1: ...42iQHL Instruction Manual High Level NO NO2 NOx Analyzer 119018 00 27Sep2018 ...
Страница 2: ......
Страница 18: ......
Страница 28: ......
Страница 36: ...Operation Instrument Display 3 8 42iQHL Instruction Manual Thermo Scientific ...
Страница 98: ...Operation Settings 3 70 42iQHL Instruction Manual Thermo Scientific Digital I O Digital I O serial number ...
Страница 100: ...Operation Settings 3 72 42iQHL Instruction Manual Thermo Scientific Sample Valve Zero Valve Span Valve Bypass Pump ...
Страница 166: ......
Страница 192: ......
Страница 198: ...Maintenance Pump Rebuilding 5 6 42iQHL Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 214: ...Servicing Fan Replacement 7 8 42iQHL Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 257: ...Servicing DMC Ozonator Thermo Scientific 42iQHL Instruction Manual 7 51 Figure 7 52 Remove Ammonia Scrubber ...
Страница 272: ......
Страница 308: ......
Страница 320: ......
Страница 321: ......