Servicing
Fuse Replacement
Thermo Scientific
42iQHL Instruction Manual
7-5
Use the following procedure to replace the fuses.
1.
Turn instrument OFF and unplug the power cord.
2.
Remove fuse drawer, located on the AC power connector.
3.
If either fuse is blown, replace both fuses.
4.
Insert fuse drawer and reconnect power cord.
Figure 7–4.
Replacing the Fuses
Fuse
Replacement