Servicing
Optional DMC Oxygen Sensor
Thermo Scientific
42iQHL Instruction Manual
7-61
Use the following to replace the oxygen sensor capillary.
Equipment required:
Phillips drive, #2
1.
Remove the oxygen sensor DMC from the instrument. Refer to
“
Oxygen Sensor Removal
” on page
7-56
.
2.
Unlock tube clamp.
3.
Unplug tube after unlocking tube clamp.
4.
Using a #2 Phillips drive, unfasten four #6-32 flat head screws.
Figure 7–64.
Replacing the Oxygen Sensor Capillary
5.
Pull capillary assembly outwards and apart as shown.
Oxygen Sensor
Capillary
Replacement