Calibration
Calibration in Dual Range and Auto Range Mode
4-16
42iQHL Instruction Manual
Thermo Scientific
The dual/auto range calibration feature is used to calibrate the analyzer at
two different span levels (as opposed to a single span level in the standard
mode) generating a “tailored multi-point” calibration curve stored in the
analyzer's memory. This feature may be used:
●
When widely different gas levels are being monitored, such as a factor
of 10 or greater apart
●
If precision and span levels are being introduced using separate tanks
●
If more than one multi-component cylinder is being used to calibrate
the instrument
Properly designed chemiluminescence analyzers are inherently linear over a
wide dynamic range; and under normal U.S. EPA compliance situations
this feature is not required. Dual calibration may be used for span levels less
than a factor of 10 apart, however if this is done to correct for a significant
non-linearity, it may mask the problems causing the effect, such as, bad
calibration cylinder, leaks in sampling lines, or low ozonator output.
Use the following procedure to set the NO and NO
x
backgrounds to zero.
Both the NO and NO
x
background screens operate the same way.
For detailed information about the menu parameters and the icons used in
these procedures, see the “
Operation
” chapter.
1.
Follow the “
Pre-Calibration
” procedure described previously in this
chapter.
2.
Be sure the instrument is in the auto or dual mode. If the instrument is
not in auto or dual mode:
a.
From the Home screen choose
Settings>Measurement
Settings>Range Mode Selection
.
b.
Select
Auto
or
Dual
mode.
3.
Introduce zero air to the SAMPLE bulkhead and allow the analyzer to
sample zero air until the NO, NO
2
, and NO
x
responses stabilize.
4.
After the responses have stabilized, from the Home Screen, choose
Calibration>Calibrate Backgrounds>Calibrate NO Background
.
Calibration in Dual
Range and Auto
Range Mode
Set NO and NO
x
Background to Zero
Содержание 42iQHL
Страница 1: ...42iQHL Instruction Manual High Level NO NO2 NOx Analyzer 119018 00 27Sep2018 ...
Страница 2: ......
Страница 18: ......
Страница 28: ......
Страница 36: ...Operation Instrument Display 3 8 42iQHL Instruction Manual Thermo Scientific ...
Страница 98: ...Operation Settings 3 70 42iQHL Instruction Manual Thermo Scientific Digital I O Digital I O serial number ...
Страница 100: ...Operation Settings 3 72 42iQHL Instruction Manual Thermo Scientific Sample Valve Zero Valve Span Valve Bypass Pump ...
Страница 166: ......
Страница 192: ......
Страница 198: ...Maintenance Pump Rebuilding 5 6 42iQHL Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 214: ...Servicing Fan Replacement 7 8 42iQHL Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 257: ...Servicing DMC Ozonator Thermo Scientific 42iQHL Instruction Manual 7 51 Figure 7 52 Remove Ammonia Scrubber ...
Страница 272: ......
Страница 308: ......
Страница 320: ......
Страница 321: ......