![Thermo Scientific 42iQHL Скачать руководство пользователя страница 173](http://html.mh-extra.com/html/thermo-scientific/42iqhl/42iqhl_instruction-manual_1105144173.webp)
Calibration
Equipment Required
Thermo Scientific
42iQHL Instruction Manual
4-7
The O
3
generator (ozonator) airflow rate and the NO flow rate must be
adjusted such that the following relationships hold:
min
-
ppm
2.75
t
x
]
[NO
=
P
R
RC
R
)
F
+
F
(
F
]
[NO
=
]
[NO
NO
O
NO
STD
RC
min
2
<
F
+
F
V
=
t
NO
O
RC
R
Use the following procedure to determine the flow conditions to be used in
the GPT system.
1.
Determine FT, the total flow required at the output manifold, which
should be equal to the analyzer demand plus 10 to 50 percent excess.
2.
Establish [NO]
OUT
as the highest NO concentration that will be
required at the output manifold. [NO]
OUT
should be about equal to
90% of the upper range limit (URL) of the NO
2
concentration range to
be covered.
3.
Determine F
NO
as:
]
[NO
F
x
]
[NO
=
F
STD
T
OUT
NO
4.
Select a convenient or available reaction chamber volume. Initially a
trial volume may be selected in the range of 200 to 500 cc.
5.
Compute F
O
as:
F
-
2.75
V
x
F
x
]
[NO
=
F
NO
RC
NO
STD
O
6.
Compute t
R
as:
Determining GPT System
Flow Conditions
Содержание 42iQHL
Страница 1: ...42iQHL Instruction Manual High Level NO NO2 NOx Analyzer 119018 00 27Sep2018 ...
Страница 2: ......
Страница 18: ......
Страница 28: ......
Страница 36: ...Operation Instrument Display 3 8 42iQHL Instruction Manual Thermo Scientific ...
Страница 98: ...Operation Settings 3 70 42iQHL Instruction Manual Thermo Scientific Digital I O Digital I O serial number ...
Страница 100: ...Operation Settings 3 72 42iQHL Instruction Manual Thermo Scientific Sample Valve Zero Valve Span Valve Bypass Pump ...
Страница 166: ......
Страница 192: ......
Страница 198: ...Maintenance Pump Rebuilding 5 6 42iQHL Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 214: ...Servicing Fan Replacement 7 8 42iQHL Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 257: ...Servicing DMC Ozonator Thermo Scientific 42iQHL Instruction Manual 7 51 Figure 7 52 Remove Ammonia Scrubber ...
Страница 272: ......
Страница 308: ......
Страница 320: ......
Страница 321: ......