Operation
Calibration
3-30
42iQHL Instruction Manual
Thermo Scientific
The Manual O
2
Calibration screen allows the O
2
span coefficients to be
changed manually while sampling span gas of known concentration. This
button appears if the O
2
Sensor option is selected in the Configuration
screen.
Home Screen>Calibration>Advanced Calibration>Manual Calibration
(single range mode)
The Manual O
2
Calibration screen contains the following information:
●
Adjust O
2
Span Coefficient:
Allows the user to manually adjust the O
2
span coefficient.
●
Reset O
2
Coefficient to 1.000:
Resets coefficient.
Manual O
2
Calibration
Содержание 42iQHL
Страница 1: ...42iQHL Instruction Manual High Level NO NO2 NOx Analyzer 119018 00 27Sep2018 ...
Страница 2: ......
Страница 18: ......
Страница 28: ......
Страница 36: ...Operation Instrument Display 3 8 42iQHL Instruction Manual Thermo Scientific ...
Страница 98: ...Operation Settings 3 70 42iQHL Instruction Manual Thermo Scientific Digital I O Digital I O serial number ...
Страница 100: ...Operation Settings 3 72 42iQHL Instruction Manual Thermo Scientific Sample Valve Zero Valve Span Valve Bypass Pump ...
Страница 166: ......
Страница 192: ......
Страница 198: ...Maintenance Pump Rebuilding 5 6 42iQHL Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 214: ...Servicing Fan Replacement 7 8 42iQHL Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 257: ...Servicing DMC Ozonator Thermo Scientific 42iQHL Instruction Manual 7 51 Figure 7 52 Remove Ammonia Scrubber ...
Страница 272: ......
Страница 308: ......
Страница 320: ......
Страница 321: ......