Calibration
Calibration in Dual Range and Auto Range Mode
Thermo Scientific
42iQHL Instruction Manual
4-17
The Target Concentration button will read 0.00. The New
Background button will display the background needed to make the
current NO concentration go to 0.00.
5.
In the Calibrate Backgrounds screen, press
Calibrate
to set the NO
reading to zero and to save the new background.
6.
Press the Back button to return to the Calibration Backgrounds screen
and repeat this procedure to set the
Calibrate NO
x
Background
to
zero.
Use the following procedure to calibrate NO channel to the NO
calibration gas.
1.
Disconnect the source of zero air from the SAMPLE bulkhead. In its
place, connect a source of NO calibration gas of about 80% of the low
NO full-scale range.
2.
Allow the analyzer to sample the low NO calibration gas until the NO,
NO
2
, and NO
x
readings stabilize.
3.
When the responses stabilize, from the Home screen, choose
Calibration>Calibrate Span Coefficients>Calibrate Low Range NO
Span Coefficient
.
The user sets the span concentration by pressing the Edit Span
Concentration button. The New Span Coefficient button will show
what the span coefficient will be set to if the Calibrate button is
pressed.
4.
Enter the NO calibration gas concentration using the pushbuttons, and
then press
Calibrate
to calculate and save the new low range NO span
coefficient based on the entered span coefficient.
Use the following procedure to calibrate the NO
x
channel to the NO
x
calibration gas.
1.
Press the Back button to return to the Calibration menu and choose
Calibrate Low Range NO
x
Span Coefficient
.
Calibrate NO Low
Calibrate NO
x
Low
Содержание 42iQHL
Страница 1: ...42iQHL Instruction Manual High Level NO NO2 NOx Analyzer 119018 00 27Sep2018 ...
Страница 2: ......
Страница 18: ......
Страница 28: ......
Страница 36: ...Operation Instrument Display 3 8 42iQHL Instruction Manual Thermo Scientific ...
Страница 98: ...Operation Settings 3 70 42iQHL Instruction Manual Thermo Scientific Digital I O Digital I O serial number ...
Страница 100: ...Operation Settings 3 72 42iQHL Instruction Manual Thermo Scientific Sample Valve Zero Valve Span Valve Bypass Pump ...
Страница 166: ......
Страница 192: ......
Страница 198: ...Maintenance Pump Rebuilding 5 6 42iQHL Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 214: ...Servicing Fan Replacement 7 8 42iQHL Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 257: ...Servicing DMC Ozonator Thermo Scientific 42iQHL Instruction Manual 7 51 Figure 7 52 Remove Ammonia Scrubber ...
Страница 272: ......
Страница 308: ......
Страница 320: ......
Страница 321: ......