Calibration
Calibration
Thermo Scientific
42iQHL Instruction Manual
4-15
11.
Record the NO
2
concentration and the analyzer's NO
2
response.
12.
Maintaining the same F
NO
, F
O
, and F
D
, adjust the ozone generator to
obtain several other concentrations of NO
2
over the NO
2
range (at least
five evenly spaced points across the remaining scale are suggested).
13.
Record the stable responses and plot the analyzer's NO
2
responses
versus the corresponding calculated (using the above equation for
[NO
2
]
OUT
) concentrations and draw or calculate the NO
2
calibration
curve.
Note
It is important that the curve be linear within
1% FS over the NO
2
range. If the curve is nonlinear, the analyzer is not operating correctly,
(possible leak, or converter failure, etc.), and should be serviced. Assuming
the curve is linear, subsequent data should be reduced using this NO
2
calibration curve response.
▲
Note
The user can also manually change the calibration factors. This is
often useful in a troubleshooting situation. From the Home screen, choose
Calibration>Advanced Calibration>Manual Calibration>
Adjust Span
Coefficients
.
However, after the above calibration procedure is completed, all subsequent
data reduction depends on the calibration parameters remaining the same
as during the initial calibration. Therefore, never change any calibration
factor without first recording the value so that after any troubleshooting
procedure is completed, the initial value can be re-entered thereby not
altering the multipoint calibration.
▲
Содержание 42iQHL
Страница 1: ...42iQHL Instruction Manual High Level NO NO2 NOx Analyzer 119018 00 27Sep2018 ...
Страница 2: ......
Страница 18: ......
Страница 28: ......
Страница 36: ...Operation Instrument Display 3 8 42iQHL Instruction Manual Thermo Scientific ...
Страница 98: ...Operation Settings 3 70 42iQHL Instruction Manual Thermo Scientific Digital I O Digital I O serial number ...
Страница 100: ...Operation Settings 3 72 42iQHL Instruction Manual Thermo Scientific Sample Valve Zero Valve Span Valve Bypass Pump ...
Страница 166: ......
Страница 192: ......
Страница 198: ...Maintenance Pump Rebuilding 5 6 42iQHL Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 214: ...Servicing Fan Replacement 7 8 42iQHL Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 257: ...Servicing DMC Ozonator Thermo Scientific 42iQHL Instruction Manual 7 51 Figure 7 52 Remove Ammonia Scrubber ...
Страница 272: ......
Страница 308: ......
Страница 320: ......
Страница 321: ......