Servicing
DMC Converter Replacement
Thermo Scientific
42iQHL Instruction Manual
7-41
Use the following procedure to replace the converter.
Equipment Required:
Phillips drive, #2
1.
Turn the instrument OFF, unplug the power cord, and remove the
cover.
2.
Unplug DMC cables 2X.
3.
Disconnect all plumbing 7X.
Note
The amount of connections will vary depending on the different
options installed.
▲
4.
Using a #2 Phillips drive, unfasten captive hardware 4X.
Figure 7–43.
Converter DMC Assembly Remove
DMC Converter
Replacement
Содержание 42iQHL
Страница 1: ...42iQHL Instruction Manual High Level NO NO2 NOx Analyzer 119018 00 27Sep2018 ...
Страница 2: ......
Страница 18: ......
Страница 28: ......
Страница 36: ...Operation Instrument Display 3 8 42iQHL Instruction Manual Thermo Scientific ...
Страница 98: ...Operation Settings 3 70 42iQHL Instruction Manual Thermo Scientific Digital I O Digital I O serial number ...
Страница 100: ...Operation Settings 3 72 42iQHL Instruction Manual Thermo Scientific Sample Valve Zero Valve Span Valve Bypass Pump ...
Страница 166: ......
Страница 192: ......
Страница 198: ...Maintenance Pump Rebuilding 5 6 42iQHL Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 214: ...Servicing Fan Replacement 7 8 42iQHL Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 257: ...Servicing DMC Ozonator Thermo Scientific 42iQHL Instruction Manual 7 51 Figure 7 52 Remove Ammonia Scrubber ...
Страница 272: ......
Страница 308: ......
Страница 320: ......
Страница 321: ......