System Description
Reaction Chamber DMC
8-2
42iQHL Instruction Manual
Thermo Scientific
●
Flow Pressure DMC with restricting capillary
●
Firmware
●
Oxygen sensor (optional)
The reaction chamber is where the sample reacts with ozone and produces
excited NO
2
that gives off a photon of energy when it decays.
The reaction chamber is heated and controlled to approximately 50 °C in
order to ensure the greatest instrument stability. The sample and ozone
flow capillaries and a thermistor sensor are also housed in/on the reaction
chamber assembly.
The optical filter housed in the reaction chamber limits the spectral region
viewed by the detector and eliminates possible interferences due to other
chemiluminescent reactions.
The Photomultiplier tube (PMT) provides the infrared sensitivity required
to detect the NO
2
luminescence resulting from the reaction of the ozone
with the ambient air sample.
Optical energy from the reaction is converted to an electrical signal by the
PMT and sent to the input board that transmits it to the processor.
The thermoelectric PMT cooler reduces the PMT temperature to
approximately -3 °C to minimize dark current and increase instrument
sensitivity. The cooler helps to increase zero and span stability over a wide
ambient temperature range. The cooler housing also shields the PMT from
external electrical and optical interferences.
The Ozonator generates the necessary ozone concentration required for the
chemiluminescent reaction. The ozone reacts with the NO in the ambient
air sample to produce the electronically excited NO
2
molecules.
The ozonator flow switch located at the ozonator inlet completes an
electrical safety circuit when air flows through the sensor to the ozonator. If
airflow stops, the flow sensor breaks the electrical circuit to the ozonator
and shuts it off to prevent the ozonator from overheating.
Reaction
Chamber DMC
Optical Filter
Photomultiplier
Tube
Photomultiplier
Tube Cooler
Ozonator
Ozonator Flow
Switch
Содержание 42iQHL
Страница 1: ...42iQHL Instruction Manual High Level NO NO2 NOx Analyzer 119018 00 27Sep2018 ...
Страница 2: ......
Страница 18: ......
Страница 28: ......
Страница 36: ...Operation Instrument Display 3 8 42iQHL Instruction Manual Thermo Scientific ...
Страница 98: ...Operation Settings 3 70 42iQHL Instruction Manual Thermo Scientific Digital I O Digital I O serial number ...
Страница 100: ...Operation Settings 3 72 42iQHL Instruction Manual Thermo Scientific Sample Valve Zero Valve Span Valve Bypass Pump ...
Страница 166: ......
Страница 192: ......
Страница 198: ...Maintenance Pump Rebuilding 5 6 42iQHL Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 214: ...Servicing Fan Replacement 7 8 42iQHL Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 257: ...Servicing DMC Ozonator Thermo Scientific 42iQHL Instruction Manual 7 51 Figure 7 52 Remove Ammonia Scrubber ...
Страница 272: ......
Страница 308: ......
Страница 320: ......
Страница 321: ......