Operation
Calibration
3-20
42iQHL Instruction Manual
Thermo Scientific
The Calibrate O
2
Span Coefficient screen is used to enter the O
2
span
concentration and calibrate the O
2
span coefficient while sampling span gas
of known concentration. This button appears if the O
2
Sensor option is
selected in the configuration screen.
It is important to note the O
2
averaging time when calibrating. The longer
the averaging time the more precise the calibration results. To achieve
maximum precision, allow the instrument to stabilize each time input gas is
changed and set the averaging time to 300-second averaging. The O
2
Averaging Time is located at Settings>Measurement Settings>Averaging
Time.
Home Screen>Calibration>Calibrate O
2
Span Coefficient (single range
mode)
The Calibrate O
2
Span Coefficients screen contains the following
information:
●
Edit O
2
Span Concentration:
User enters the O
2
span concentration.
●
Current O
2
Concentration:
Read only. Current O
2
concentration
reading.
●
Current O
2
Span Coefficient:
Read only. Current O
2
span coefficient
value.
●
Calculated O
2
Span Coefficient:
Read only. After the “Edit O
2
Span
Concentration” value is entered, the new calculated O
2
span coefficient
is displayed.
●
Calibrate:
When pressed, updates the O
2
coefficient and the O
2
concentration should match the span concentration.
Calibrate O
2
Span
Coefficient
Содержание 42iQHL
Страница 1: ...42iQHL Instruction Manual High Level NO NO2 NOx Analyzer 119018 00 27Sep2018 ...
Страница 2: ......
Страница 18: ......
Страница 28: ......
Страница 36: ...Operation Instrument Display 3 8 42iQHL Instruction Manual Thermo Scientific ...
Страница 98: ...Operation Settings 3 70 42iQHL Instruction Manual Thermo Scientific Digital I O Digital I O serial number ...
Страница 100: ...Operation Settings 3 72 42iQHL Instruction Manual Thermo Scientific Sample Valve Zero Valve Span Valve Bypass Pump ...
Страница 166: ......
Страница 192: ......
Страница 198: ...Maintenance Pump Rebuilding 5 6 42iQHL Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 214: ...Servicing Fan Replacement 7 8 42iQHL Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 257: ...Servicing DMC Ozonator Thermo Scientific 42iQHL Instruction Manual 7 51 Figure 7 52 Remove Ammonia Scrubber ...
Страница 272: ......
Страница 308: ......
Страница 320: ......
Страница 321: ......