![Thermo Scientific 42iQHL Скачать руководство пользователя страница 179](http://html.mh-extra.com/html/thermo-scientific/42iqhl/42iqhl_instruction-manual_1105144179.webp)
Calibration
Calibration
Thermo Scientific
42iQHL Instruction Manual
4-13
is pressed. Pressing the Calibrate button will save the new NO
x
span
coefficient and calibrate the instrument.
3.
Enter the NO
x
calibration gas concentration using the pushbuttons,
and then press
Calibrate
to calibrate the instrument to the NO
x
calibration gas.
The exact NO
x
concentration is calculated from:
F
+
F
+
F
)
]
NO
[
+
]
[NO
(
x
F
=
]
NO
[
D
O
NO
IMP
2
STD
NO
OUT
x
Where:
[NO
x
]
OUT
= diluted NO
x
concentration at the output manifold, ppm
[NO
2
]
IMP
= concentration of NO
2
impurity in the standard NO cylinder, ppm
4.
Record the NO
x
concentration and the analyzer's NO
x
response.
Use the following procedures to prepare the NO, NO
x
, and NO
2
calibration curves.
1.
Generate several additional NO and NO
x
concentrations by decreasing
F
NO
or increasing F
D
.
2.
For each concentration generated, calculate the exact NO and NO
x
concentrations using the above equations for [NO]
OUT
and [NO
x
]
OUT
.
3.
Record the NO and NO
x
responses.
4.
Plot the analyzer responses versus the respective calculated NO and
NO
x
concentrations and draw or calculate the respective calibration
curves. For subsequent calibrations where linearity can be assumed,
these curves may be checked with a three-point calibration consisting of
a zero point, NO and NO
x
concentrations of approximately 80% of the
URL, and an intermediate concentration.
5.
Adjust the GPT system to generate a NO concentration near 90% of
the URL of the instrument range selected.
6.
Sample this NO concentration until the NO and NO
x
responses have
stabilized, then measure and record the NO concentration as [NO]
ORIG
.
Preparing NO, NO
x
, and
NO
2
Calibration Curves
Содержание 42iQHL
Страница 1: ...42iQHL Instruction Manual High Level NO NO2 NOx Analyzer 119018 00 27Sep2018 ...
Страница 2: ......
Страница 18: ......
Страница 28: ......
Страница 36: ...Operation Instrument Display 3 8 42iQHL Instruction Manual Thermo Scientific ...
Страница 98: ...Operation Settings 3 70 42iQHL Instruction Manual Thermo Scientific Digital I O Digital I O serial number ...
Страница 100: ...Operation Settings 3 72 42iQHL Instruction Manual Thermo Scientific Sample Valve Zero Valve Span Valve Bypass Pump ...
Страница 166: ......
Страница 192: ......
Страница 198: ...Maintenance Pump Rebuilding 5 6 42iQHL Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 214: ...Servicing Fan Replacement 7 8 42iQHL Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 257: ...Servicing DMC Ozonator Thermo Scientific 42iQHL Instruction Manual 7 51 Figure 7 52 Remove Ammonia Scrubber ...
Страница 272: ......
Страница 308: ......
Страница 320: ......
Страница 321: ......