![Thermo Scientific 42iQHL Скачать руководство пользователя страница 269](http://html.mh-extra.com/html/thermo-scientific/42iqhl/42iqhl_instruction-manual_1105144269.webp)
Servicing
Optional DMC Oxygen Sensor
Thermo Scientific
42iQHL Instruction Manual
7-63
Use the following to replace the oxygen sensor.
Equipment required:
Hex drive, 3mm
1.
Remove the oxygen sensor DMC from the instrument. Refer to
“
Oxygen Sensor Removal
” on page
7-56
.
2.
Unplug oxygen sensor cable.
Figure 7–66.
Replacing the Oxygen Sensor
3.
Disconnect tubing.
4.
Using a 3mm hex drive, unfasten three m4 socket cap head screws.
5.
Slide out oxygen sensor.
6.
Replace oxygen sensor and reassemble in reverse order.
Oxygen Sensor
Replacement
Содержание 42iQHL
Страница 1: ...42iQHL Instruction Manual High Level NO NO2 NOx Analyzer 119018 00 27Sep2018 ...
Страница 2: ......
Страница 18: ......
Страница 28: ......
Страница 36: ...Operation Instrument Display 3 8 42iQHL Instruction Manual Thermo Scientific ...
Страница 98: ...Operation Settings 3 70 42iQHL Instruction Manual Thermo Scientific Digital I O Digital I O serial number ...
Страница 100: ...Operation Settings 3 72 42iQHL Instruction Manual Thermo Scientific Sample Valve Zero Valve Span Valve Bypass Pump ...
Страница 166: ......
Страница 192: ......
Страница 198: ...Maintenance Pump Rebuilding 5 6 42iQHL Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 214: ...Servicing Fan Replacement 7 8 42iQHL Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 257: ...Servicing DMC Ozonator Thermo Scientific 42iQHL Instruction Manual 7 51 Figure 7 52 Remove Ammonia Scrubber ...
Страница 272: ......
Страница 308: ......
Страница 320: ......
Страница 321: ......