Servicing
Measurement Side Removal and Replacing
Thermo Scientific
42iQHL Instruction Manual
7-11
6.
Swing arm open.
7.
Unfasten captive hardware.
8.
Remove two 8-32 flat head screws.
9.
Pull measurement side out.
10.
Replace and reassemble in reverse order.
Figure 7–12.
Unfasten Hardware Rear for Measurement Side Removal