Operation
Calibration
3-34
42iQHL Instruction Manual
Thermo Scientific
The Calibration History screen shows the log of calibrations and
calibration checks performed.
Use the
and
buttons to move up and down and the
and
buttons to move left and right.
Home Screen>Calibration>Advanced Calibration>Calibration History
Note
Pressing the Calibration History button responds with Retrieving
calibration log data, it may take a few seconds...
▲
The Calibration History screen contains the following information:
●
Time Stamp:
Time of calibration or calibration check.
●
Event:
Lists the type of calibration event.
●
Result:
Concentration result.
●
Target:
Concentration setpoint value.
●
Units:
Displays units for each item.
●
Average Time:
Averaging time used during the calibration or calibration
check.
Calibration History
Содержание 42iQHL
Страница 1: ...42iQHL Instruction Manual High Level NO NO2 NOx Analyzer 119018 00 27Sep2018 ...
Страница 2: ......
Страница 18: ......
Страница 28: ......
Страница 36: ...Operation Instrument Display 3 8 42iQHL Instruction Manual Thermo Scientific ...
Страница 98: ...Operation Settings 3 70 42iQHL Instruction Manual Thermo Scientific Digital I O Digital I O serial number ...
Страница 100: ...Operation Settings 3 72 42iQHL Instruction Manual Thermo Scientific Sample Valve Zero Valve Span Valve Bypass Pump ...
Страница 166: ......
Страница 192: ......
Страница 198: ...Maintenance Pump Rebuilding 5 6 42iQHL Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 214: ...Servicing Fan Replacement 7 8 42iQHL Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 257: ...Servicing DMC Ozonator Thermo Scientific 42iQHL Instruction Manual 7 51 Figure 7 52 Remove Ammonia Scrubber ...
Страница 272: ......
Страница 308: ......
Страница 320: ......
Страница 321: ......