![Thermo Scientific 42iQHL Скачать руководство пользователя страница 125](http://html.mh-extra.com/html/thermo-scientific/42iqhl/42iqhl_instruction-manual_1105144125.webp)
Operation
Settings
Thermo Scientific
42iQHL Instruction Manual
3-97
The Compensation screen provides compensation for any changes to the
instrument’s output signal due to internal instrument temperature,
pressure, and oxygen variations (if enabled).
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings>Compensation
The Compensation screen contains the following information:
●
Temp Compensation:
Toggles temperature compensation enabled or
disabled and provides compensation for any changes to the
instrument's output signal due to internal instrument temperature
variations. The effects of internal instrument temperature changes on
the analyzer's subsystems and output have been empirically determined.
This empirical data is used to compensate for any changes in
temperature.
●
Pressure Compensation:
Toggles pressure compensation enabled or
disabled and provides compensation for any changes to the
instrument's output signal due to bench pressure variations. The effects
of bench pressure changes on the analyzer's subsystems and output have
been empirically determined. This empirical data is used to compensate
for any change in chamber pressure.
●
O
2
Correction:
Allows for the correction of the NO, NO
2
, and NO
x
concentrations based on O
2
readings.
●
Standard Temperature for Gas Units:
Allows the user to enter a value
between 0 and 30
°
C to correct the
g/m
3
, mg/m
3
, and g/m
3
concentration calculations.
Compensation
Содержание 42iQHL
Страница 1: ...42iQHL Instruction Manual High Level NO NO2 NOx Analyzer 119018 00 27Sep2018 ...
Страница 2: ......
Страница 18: ......
Страница 28: ......
Страница 36: ...Operation Instrument Display 3 8 42iQHL Instruction Manual Thermo Scientific ...
Страница 98: ...Operation Settings 3 70 42iQHL Instruction Manual Thermo Scientific Digital I O Digital I O serial number ...
Страница 100: ...Operation Settings 3 72 42iQHL Instruction Manual Thermo Scientific Sample Valve Zero Valve Span Valve Bypass Pump ...
Страница 166: ......
Страница 192: ......
Страница 198: ...Maintenance Pump Rebuilding 5 6 42iQHL Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 214: ...Servicing Fan Replacement 7 8 42iQHL Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 257: ...Servicing DMC Ozonator Thermo Scientific 42iQHL Instruction Manual 7 51 Figure 7 52 Remove Ammonia Scrubber ...
Страница 272: ......
Страница 308: ......
Страница 320: ......
Страница 321: ......