Servicing
DMC PMT Cooler and Reaction Chamber Replacement
Thermo Scientific
42iQHL Instruction Manual
7-35
Use the following procedure to clean or remove the reaction chamber.
Equipment Required:
Phillips drive, #2
1.
Turn the instrument OFF, unplug the power cord, and remove the
cover.
2.
Remove DMC Converter as described in “
PMT Cooler Shroud
Removal
” in this chapter.
3.
Unplug heater J8 from board.
Figure 7–36.
Photo Reaction Chamber Cleaning, Electrical
4.
Disconnect plumbing as shown.
5.
Using a 9/64-inch hex drive, unfasten three #8-32 socket cap head
screws.
Reaction Chamber
Cleaning and/or
Removal
Содержание 42iQHL
Страница 1: ...42iQHL Instruction Manual High Level NO NO2 NOx Analyzer 119018 00 27Sep2018 ...
Страница 2: ......
Страница 18: ......
Страница 28: ......
Страница 36: ...Operation Instrument Display 3 8 42iQHL Instruction Manual Thermo Scientific ...
Страница 98: ...Operation Settings 3 70 42iQHL Instruction Manual Thermo Scientific Digital I O Digital I O serial number ...
Страница 100: ...Operation Settings 3 72 42iQHL Instruction Manual Thermo Scientific Sample Valve Zero Valve Span Valve Bypass Pump ...
Страница 166: ......
Страница 192: ......
Страница 198: ...Maintenance Pump Rebuilding 5 6 42iQHL Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 214: ...Servicing Fan Replacement 7 8 42iQHL Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 257: ...Servicing DMC Ozonator Thermo Scientific 42iQHL Instruction Manual 7 51 Figure 7 52 Remove Ammonia Scrubber ...
Страница 272: ......
Страница 308: ......
Страница 320: ......
Страница 321: ......