Troubleshooting
Troubleshooting Guide
6-6
42iQHL Instruction Manual
Thermo Scientific
Problem
Possible Cause
Action
Alarm – Cooler Voltage too
High
Defective cooler
Replace cooler as needed.
Defective DMC board
Replace DMC board as needed.
Alarm – PMT Voltage too
Low
Defective PMT
Replace PMT as needed.
Defective DMC board
Replace DMC board as needed.
Alarm – PMT Voltage too
High
Defective PMT
Replace PMT as needed.
Defective DMC board
Replace DMC board as needed.
Alarm – Frequency too Low
Input board is out of
calibration
Recalibrate the Input board.
Alarm – Frequency too High
Input board is out of
calibration
Recalibrate the Input board.
Содержание 42iQHL
Страница 1: ...42iQHL Instruction Manual High Level NO NO2 NOx Analyzer 119018 00 27Sep2018 ...
Страница 2: ......
Страница 18: ......
Страница 28: ......
Страница 36: ...Operation Instrument Display 3 8 42iQHL Instruction Manual Thermo Scientific ...
Страница 98: ...Operation Settings 3 70 42iQHL Instruction Manual Thermo Scientific Digital I O Digital I O serial number ...
Страница 100: ...Operation Settings 3 72 42iQHL Instruction Manual Thermo Scientific Sample Valve Zero Valve Span Valve Bypass Pump ...
Страница 166: ......
Страница 192: ......
Страница 198: ...Maintenance Pump Rebuilding 5 6 42iQHL Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 214: ...Servicing Fan Replacement 7 8 42iQHL Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 257: ...Servicing DMC Ozonator Thermo Scientific 42iQHL Instruction Manual 7 51 Figure 7 52 Remove Ammonia Scrubber ...
Страница 272: ......
Страница 308: ......
Страница 320: ......
Страница 321: ......