![Thermo Scientific 42iQD Скачать руководство пользователя страница 218](http://html.mh-extra.com/html/thermo-scientific/42iqd/42iqd_instruction-manual_1105143218.webp)
Servicing
DMC PMT Cooler and Reaction Chamber Replacement
7-28
42iQD Instruction Manual
Thermo Scientific
Figure 7–32.
Remove PMT Cooler
Use the following procedure to replace the PMT cooler board.
Equipment Required:
Phillips drive, #2
1.
Turn the instrument OFF, unplug the power cord, and remove the
cover.
2.
Remove DMC Converter as described in “
” in this chapter.
3.
Unplug all electrical connections from the board.
4.
Using a #2 Phillips drive, unfasten six #6-32 pan head screws.
5.
Pull board outwards.
PMT Cooler Board
Replacement