Servicing
Pump Replacement
Thermo Scientific
42iQD Instruction Manual
7-19
Use the following procedure to replace the pump.
Equipment required:
Phillips drive, #1 and #2
1.
Turn instrument OFF, unplug power cord, and remove the cover
2.
Unplug pump cable from Step POL board J7.
3.
Disconnect tubing from pump.
4.
Unfasten two captive hardware.
5.
Slide pump right until keyway meets opening.
Figure 7–22.
Remove Pump, Disconnect and Unfasten
6.
Pull pump outwards.
Note
When installing pump, make sure the pump keyway opening goes
over the keyway.
▲
Pump
Replacement