![Thermo Scientific 42iQD Скачать руководство пользователя страница 9](http://html.mh-extra.com/html/thermo-scientific/42iqd/42iqd_instruction-manual_1105143009.webp)
Thermo Scientific
42iQD Instruction Manual
1-1
Chapter 1
Introduction
The Thermo Scientific™ 42iQD NO
x
Analyzer utilizes chemiluminescence
technology to measure the amount of nitrogen oxides in the air from sub-
ppb levels up to 100 ppm.
This analyzer is a single chamber, single photomultiplier tube design that
measures NO
x
only. Dual range and auto range are standard features in this
instrument. The instrument operates continuously in the NO
x
mode,
providing a response time of less than 10 seconds.
Temperature and pressure correction are standard offerings of the 42iQD
analyzer. User settable alarm levels for concentration and a variety of
internal diagnostics are available from an easy to follow menu structure.
The iQ Series Instrument Platform is a smart environmental monitoring
solution for ambient and source gas analysis that affords greater control
over instrument performance and data availability.
●
Distributed Measurement and Control (DMC) module design
simplifies serviceability. Each DMC module contains its own
microprocessor control enabling functional performance validation
at the module level.
●
Built-in predictive diagnostics and preventive maintenance
schedules identify problems before they occur. The iQ Series
platform sends email notifications directly to Thermo Fisher
Scientific’s world class service support team or locally identified
addressees in order to proactively communicate analyzer
performance conditions and identify spare parts needs before an
operational concern arises.
●
The iQ Series platform supports Modbus, streaming and VNC
protocols over serial and Ethernet as well as analog and digital I/O
for easy integration into most data management systems.
●
Three standard USB ports afford convenient data download
capability as well as the ability to connect additional hardware, such
as a computer keyboard or mouse.
●
The iQ Series GUI runs on a 7” color touchscreen display. The
GUI is highly flexible and can be customized to enable a tailored
iQ Series
Instrument
Platform
Содержание 42iQD
Страница 1: ...42iQD Instruction Manual NOx Analyzer 119013 00 1Sep2021 ...
Страница 2: ......
Страница 32: ...Operation Instrument Display 3 8 42iQD Instruction Manual Thermo Scientific ...
Страница 182: ...Maintenance Pump Rebuilding 5 8 42iQD Instruction Manual Thermo Scientific Figure 5 3 Pump Rebuilding ...
Страница 184: ......
Страница 198: ...Servicing Fan Replacement 7 8 42iQD Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 212: ...Servicing Power Supply Replacement 7 22 42iQD Instruction Manual Thermo Scientific Figure 7 26 Replacing Power Supply ...
Страница 235: ...Servicing DMC Ozonator Thermo Scientific 42iQD Instruction Manual 7 45 Figure 7 49 Flow Switch Remove ...
Страница 238: ......
Страница 272: ......
Страница 284: ......
Страница 285: ......